dc.contributor.author |
Nagaraju, K |
|
dc.contributor.author |
Sankarasubramanian, K |
|
dc.contributor.author |
Rangarajan, K. E |
|
dc.contributor.author |
Ramesh, K. B |
|
dc.contributor.author |
Singh, J |
|
dc.contributor.author |
Devendran, P |
|
dc.contributor.author |
Hariharan |
|
dc.date.accessioned |
2009-08-07T10:18:26Z |
|
dc.date.available |
2009-08-07T10:18:26Z |
|
dc.date.issued |
2008-09 |
|
dc.identifier.citation |
Bulletin of the Astronomical Society of India, Vol. 36, No. 2 & 3, pp. 99 – 110 |
en |
dc.identifier.issn |
0304-9523 |
|
dc.identifier.uri |
http://hdl.handle.net/2248/4731 |
|
dc.description.abstract |
Calibration and characterization of a dual-beam polarimeter installed at Kodaikanal Tower Telescope (KTT) for spectropolarimetric observations are presented in this paper. It was found that a slit width of 48 μ is optimum for the spectrograph setup at KTT and corresponding spectral resolution is 32.47 m\AA. It was demonstrated that the precision in polarization measurement can be achieved better than 0.1 % by increasing the exposure time. However, the polarimetric calibration accuracy is limited to 0.35% for Stokes Q and U parameters and 0.2% for Stokes V parameter, mainly due to the uncertainty in the retardance of the calibration retarder. A comparison of the magnetic field measurement between the Solar Optical Telescope (SOT) onboard Hinode and KTT for an active region was made and a good match was found after spatial smearing of Hinode/SOT measurements by 5.12''. |
en |
dc.language.iso |
en |
en |
dc.publisher |
Astronomical Society of India |
en |
dc.relation.uri |
http://www.ncra.tifr.res.in/~basi/08JunSep/BASI_ABS/36992008.htm |
en |
dc.rights |
© Astronomical Society of India |
en |
dc.subject |
Instrumentation: Polarimeter |
en |
dc.title |
On the performance of a dual-beam polarimeter at Kodaikanal tower telescope |
en |
dc.type |
Article |
en |