Please use this identifier to cite or link to this item: http://hdl.handle.net/2248/8833
Title: DMD-based multi-object spectrograph (D-MOS): AIV and first light results
Authors: Sriram, S
Valsan, Vineeth
Remya, B. S
Subramaniam, A
Maheswar, G
Keywords: Digital micromirror device
Multi-object spectroscopy
Astronomical instrumentation
UV astronomy
Future UV space mission
INSIST
Issue Date: Oct-2025
Publisher: Springer Nature
Citation: Experimental Astronomy, Vol. 60, No. 2, 16
Abstract: A Digital Micromirror Device (DMD)-based Multi-Object Spectrograph (D-MOS) with an integrated imager has been developed. The optical performance of the MOS is evaluated through comprehensive laboratory calibration and on-sky observations using the 1.3-meter J.C. Bhattacharya (JCB) Telescope at the Vainu Bappu Observatory (VBO). The system is designed to assess the viability of using a DMD as a programmable slit mechanism for future ultraviolet-optical space missions. A complete imager-cum-spectrograph assembly was constructed using off-the-shelf optical components and configured for operation in the optical band, employing a DLP9500 DMD with a 1920×1080 micromirror array. Calibration experiments established the DMD-to-detector coordinate mapping and validated the strategies for object selection and slit placement. On-sky tests in crowded stellar fields confirmed successful slit targeting, precise object alignment, and multiplexed spectral acquisition. The spectrograph achieved a peak efficiency of 32%, a spectral resolving power of R∼1000 at 6000Å, a multiplexing capability of up to 46 slits (extendable to 85), and a contrast ratio of ∼ 6000. These results demonstrate the robustness and effectiveness of the DMD MOS system under real observational conditions and raise its TRL level for use in next-generation spectroscopic space missions.
Description: Restricted Access
The original publication is available at springerlink.com
URI: http://hdl.handle.net/2248/8833
ISSN: 0922-6435
Appears in Collections:IIAP Publications

Files in This Item:
File Description SizeFormat 
DMD-based Multi-Object Spectrograph (D-MOS) AIV and first light results.pdf3.41 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.